description abstractResponse functions of a 3×3 in. NaI detector, which is mainly used in PGNAA

applications, have been calculated by using MCNP-4C code. Calculated results are compared

with measured data by using standard γ-ray sources to check their accuracy. γ-rays from

radioisotope sources were used in the range from 0.081 to 4.438 MeV for this determination.

Through the precise modeling of the detector structure, the agreement between both results

has been improved.
en
languageEnglish
titleResponse Function of a 3×3 in. NaI Scintillation Detector in the range of 0.081 to 4.438 MeVen