Response Function of a 3×3 in. NaI Scintillation Detector in the range of 0.081 to 4.438 MeV

Author:
سیدهاشم میری حکیم آباد
,
Seyyed Hashem Miri Hakimabad
Year
: 2007
Abstract: Response functions of a 3×3 in. NaI detector, which is mainly used in PGNAA

applications, have been calculated by using MCNP-4C code. Calculated results are compared

with measured data by using standard γ-ray sources to check their accuracy. γ-rays from

radioisotope sources were used in the range from 0.081 to 4.438 MeV for this determination.

Through the precise modeling of the detector structure, the agreement between both results

has been improved.
URI: http://libsearch.um.ac.ir:80/fum/handle/fum/3369307
Keyword(s): FWHM,GEB,NaI (Tl),MCNP code,Monte Carlo simulation,Nonlinear response,

Response function
,
PGNAA
Collections :
  • ProfDoc
  • Show Full MetaData Hide Full MetaData
  • Statistics

    Response Function of a 3×3 in. NaI Scintillation Detector in the range of 0.081 to 4.438 MeV

Show full item record

contributor authorسیدهاشم میری حکیم آبادen
contributor authorSeyyed Hashem Miri Hakimabadfa
date accessioned2020-06-06T13:47:43Z
date available2020-06-06T13:47:43Z
date issued2007
identifier urihttp://libsearch.um.ac.ir:80/fum/handle/fum/3369307
description abstractResponse functions of a 3×3 in. NaI detector, which is mainly used in PGNAA

applications, have been calculated by using MCNP-4C code. Calculated results are compared

with measured data by using standard γ-ray sources to check their accuracy. γ-rays from

radioisotope sources were used in the range from 0.081 to 4.438 MeV for this determination.

Through the precise modeling of the detector structure, the agreement between both results

has been improved.
en
languageEnglish
titleResponse Function of a 3×3 in. NaI Scintillation Detector in the range of 0.081 to 4.438 MeVen
typeJournal Paper
contenttypeExternal Fulltext
subject keywordsFWHMen
subject keywordsGEBen
subject keywordsNaI (Tl)en
subject keywordsMCNP codeen
subject keywordsMonte Carlo simulationen
subject keywordsNonlinear responseen
subject keywords

Response function
en
subject keywordsPGNAAen
journal titleasian journal of experimental sciencesen
journal titleAsian Journal Of Experimental Sciencesfa
journal volume0
journal issue0
identifier linkhttps://profdoc.um.ac.ir/paper-abstract-1008609.html
identifier articleid1008609
  • About Us
نرم افزار کتابخانه دیجیتال "دی اسپیس" فارسی شده توسط یابش برای کتابخانه های ایرانی | تماس با یابش
DSpace software copyright © 2019-2022  DuraSpace