Investigation of Activation Rate Uniformity in aPrompt-γ Rays IVNAA Facility
Author:
, , ,Year
: 2010
Abstract: The preliminary design studies for developing a prompt-γ rays in vivo neutron activation analysis facility are in construction at Neutron Activation Research Centre of Ferdowsi University of Mashhad. In this work, activation rate distribution is studied in different bilateral configurations in order to reduce the CV of activation rate distribution throughout the body to less than 10%. It means that the maximum RMS deviation from the mean value of distribution must be smaller than 0.5%. The best case includes two pairs of 241Am-Be neutron sources. Each pair (with 70 cm distance between two sources) is positioned within up and down 40 × 110 cm2 area graphite collimators. Also, four moderator/reflector objects are added to transversal body surfaces with 5 cm thick water and 3 cm thick graphite. The proper thickness of polyethylene pre-moderators is chosen 8 mm. The RMS deviation of distribution is 0.3% with the mean activation rate of 1.85 × 10 –6 cm–3.
Keyword(s): Activation rate uniformity,IVNAA,MCNP,Collimator aperture,Moderator-reflector
Collections
:
-
Statistics
Investigation of Activation Rate Uniformity in aPrompt-γ Rays IVNAA Facility
Show full item record
contributor author | سیدهاشم میری حکیم آباد | en |
contributor author | لاله رفعت متولی | en |
contributor author | Seyyed Hashem Miri Hakimabad | fa |
contributor author | Laleh Rafat Motavali | fa |
date accessioned | 2020-06-06T14:24:14Z | |
date available | 2020-06-06T14:24:14Z | |
date issued | 2010 | |
identifier uri | http://libsearch.um.ac.ir:80/fum/handle/fum/3395208 | |
description abstract | The preliminary design studies for developing a prompt-γ rays in vivo neutron activation analysis facility are in construction at Neutron Activation Research Centre of Ferdowsi University of Mashhad. In this work, activation rate distribution is studied in different bilateral configurations in order to reduce the CV of activation rate distribution throughout the body to less than 10%. It means that the maximum RMS deviation from the mean value of distribution must be smaller than 0.5%. The best case includes two pairs of 241Am-Be neutron sources. Each pair (with 70 cm distance between two sources) is positioned within up and down 40 × 110 cm2 area graphite collimators. Also, four moderator/reflector objects are added to transversal body surfaces with 5 cm thick water and 3 cm thick graphite. The proper thickness of polyethylene pre-moderators is chosen 8 mm. The RMS deviation of distribution is 0.3% with the mean activation rate of 1.85 × 10 –6 cm–3. | en |
language | English | |
title | Investigation of Activation Rate Uniformity in aPrompt-γ Rays IVNAA Facility | en |
type | Journal Paper | |
contenttype | External Fulltext | |
subject keywords | Activation rate uniformity | en |
subject keywords | IVNAA | en |
subject keywords | MCNP | en |
subject keywords | Collimator aperture | en |
subject keywords | Moderator-reflector | en |
journal title | Journal of Radiation Research | fa |
pages | 123-130 | |
journal volume | 51 | |
journal issue | 2 | |
identifier link | https://profdoc.um.ac.ir/paper-abstract-1015183.html | |
identifier articleid | 1015183 |